Benchtop COD and Multiparameter Photometer for Wastewater Analysis - HI83214

More Views

Benchtop COD and Multiparameter Photometer for Wastewater Analysis - HI83214

$1,980.00
OR

Availability: Please note - All prices are inclusive of GST and not all items are stock items, if you require an immediate solution please contact us today.

The HI83214 COD Meter and Multiparameter Photometer for chemical oxygen demand (COD) and specific ion measurement combines accuracy and ease of use in an ergonomic benchtop design. This meter is one of the most versatile photometers in the market, offering 15 different measurement methods including COD using ready-made liquid or powder reagents. The HI83214 has many advanced features including PC connectivity via USB, a graphic LCD to display various chemical forms, and a step-by-step tutorial mode for inexperienced users.



parameters measured


15 measurement methods including COD


USB


USB for data export


timer


Built-in timer for measurement consistency and accuracy



view compatible reagents

Details

From ammonia to phosphorus, the HI83214 benchtop photometer offers 15 measurement methods for different key water quality parameters including chemical oxygen demand (COD) in 3 different ranges. This photometer features an advanced optical system that uses special tungsten lamps, narrow band interference filters, and silicon photodetectors to ensure accurate photometric readings every time. The HI83214 uses a graphic backlit LCD that allows for an intuitive user interface, offering a tutorial mode that gives a step-by-step procedure for performing a measurement. The result obtained can be displayed in various chemical forms based on the user’s preference. For tracking of data, results can be logged and then exported to a Windows® compatible PC using the HI92000 software and HI920013 USB cable.

 

Features at-a-glance

Backlit Graphic LCD Display - The HI83214 features an adjustable backlit graphic display with virtual keys and on-screen help to provide for an intuitive user interface.

Data Logging - Users can store up to 200 readings by simply pressing the LOG key. Logged readings are just as easily recalled by pressing the dedicated RCL button. Stored data includes parameter, test results, sample number, lot number, instrument ID, date and time.

PC Connectivity - Logged readings can be quickly and easily transferred to a PC via USB using the HI92000 Windows® compatible software.

Result Conversion - Eliminates confusion by automatically converting readings to other chemical forms. Common conversions are available at the touch of a button.

On-screen Tutorial - With the tutorial function enabled, short guides relating to the current operation are displayed. At any stage in the setup or measurement process, the dedicated HELP button is also available to show contextual help.

Built-in Timer - Display of time remaining before a measurement is taken. Ensures that all readings are taken at the appropriate reaction intervals for the test being performed.

Error Messages - Messages on display alerting to problems including no cap, high zero, and standard too low.

Cooling Lamp Indicator - To maintain the desirable wavelength to be used for absorbance, it is necessary to ensure components are not overheated from the heat generated by the tungsten lamp. Each photometer is designed to allow a minimal amount of time for components to cool. The cooling lamp indicator is displayed prior to a reading being taken.

Units of Measure - Appropriate unit of measure is displayed along with reading.


On-screen Features

multiparameter photometer method

Method Selection
Users can easily select any one of the 15 measurement methods via the dedicated METHOD button.

multiparameter logging

Data Logging
Up to 200 measurement readings can be logged and recalled for future use.

multiparameter logging recall

Log Recall
Logged data can easily be recalled and the chemical form can be converted at the touch of a button.

 
 


Advanced Optical System

The HI83214 features 15 different methods that can be measured using ready-made liquid or powder reagents. When the appropriate reagent is used for a selected method, the reaction between the reagent and the sample causes a color change; the greater the concentration, the deeper the color. The associated color change is then colorimetrically analyzed according to the Beer-Lambert Law. This principle states that light is absorbed by a complementary color, and the emitted radiation is dependent upon concentration. A narrow band interference filter allows the relevant wavelength of light to be detected by the silicon photodetector and omits all other visible light emitted from the tungsten lamp. As the change in color of the reacted sample increases, absorbance of the specific wavelength of light also increases, while transmittance decreases.  

photometer multiparameter optical
 




Parameters

Reagents are available in liquid or powder form, and the amount of each reagent is precisely dosed to ensure maximum repeatability.


Parameter

Range

Resolution

Accuracy

Method

Reagent Code

COD LR

0 to 150 mg/L

1 mg/L

±5 mg/L ±5% of reading

dichromate EPAǂ

dichromate mercury-free◊◊

dichromate ISO◊

HI93754A-25

HI93754D-25

HI93754F-25

COD MR

0 to 1500 mg/L

1 mg/L

±15 mg/L ±4% of reading

dichromate EPAǂ

dichromate mercury-free◊◊

dichromate ISO◊

HI93754B-25

HI93754E-25

HI93754G-25

COD HR

0 to 15000 mg/L

10 mg/L

±150 mg/L ±3% of reading

dichromate

HI93754C-25

ǂMethod with chromium-sulfuric acid is officially recognized by EPA for wastewater analysis.  

The HI94754F-25 and HI94754G-25 method follows the official method ISO 15705.

◊◊This method is recommended for general purpose analysis with no chloride interference.  



Parameter

Range

Resolution

Accuracy*

Method

Reagent Code

Ammonia LR

0.00 to 3.00 mg/L (ppm) as

NH3-N

0.01 mg/L

±0.10 mg/L ±5% of reading

Nessler

HI93764A-25 (25 tests)

Ammonia HR

0 to 100 mg/L (ppm) as NH3-N

1 mg/L

±1 mg/L ±5% of reading

Nessler

HI93764B-25 (25 tests)

Chlorine, Free

0.00 to 5.00 mg/L (ppm)

0.01 mg/L below 0.99 mg/L; 0.1 mg/L above 0.99 mg/L

±0.03 mg/L ±4% of reading

DPD

HI93701-01 (100 tests)

HI93701-03 (300 tests)

Chlorine, Total

0.00 to 5.00 mg/L (ppm)

0.01 mg/L below 0.99 mg/L; 0.1 mg/L above 0.99 mg/L

±0.03 mg/L ±4% of reading

DPD

HI93711-01 (100 tests)

HI93711-03 (300 tests)

Nitrate

0.0 to 30.0 mg/L (ppm) as NO3-N

0.1 mg/L

±1.0 mg/L ±5% of reading @ 20°C

chromotropic acid

HI93766-50 (50 tests)

Nitrogen, Total LR

0.0 to 25.0 mg/L (ppm)

0.1 mg/L

±1.0 mg/L ±5% of reading @ 20°C

chromotropic acid

HI93767A-50 (50 tests)

Nitrogen, Total HR

10 to 150 mg/L (ppm) as N

1 mg/L

±3 mg/L ±4% of reading

chromotropic acid

HI93767B-50 (50 tests)

Phosphorus, Acid Hydrolyzable

0.00 to 1.60 mg/L (ppm) as P

0.01 mg/L

±0.05 mg/L ±5% of reading

ascorbic acid

HI93758B-50 (50 tests)

Phosphorus, Reactive

0.00 to 1.60 mg/L (ppm) as P

0.01 mg/L

±0.05 mg/L ±5% of reading

ascorbic acid

HI93758A-50 (50 tests)

Phosphorus, Reactive HR

0.0 to 32.6 mg/L (ppm) as P

0.1 mg/L

±0.5 mg/L ±5% of reading

vanadomolybdophosphoric acid

HI93763A-50 (50 tests)

Phosphorus, Total

0.00 to 1.15 mg/L (ppm) as P

0.01 mg/L

±0.05 mg/L ±6% of reading

ascorbic acid

HI93758C-50 (50 tests)

Phosphorus, Total HR

0.0 to 32.6 mg/L (ppm)

0.1 mg/L

±0.5 mg/L ±5% of reading

vanadomolybdophosphoric acid

HI93763B-50 (50 tests)

*      @25°C (77°F) unless otherwise stated



Ammonia

Ammonia Range Low Range: 0.00 to 3.00 mg/L (as NH3-N)
High Range: 0.00 to 3.00 mg/L (ppm) (as NH3-N)
Ammonia Resolution Low Range: 0.01 mg/L
High Range: 1 mg/L
Ammonia Accuracy Low Range: ±0.10 mg/L ±5% of reading
High Range: ±1 mg/L ±5% of reading
Ammonia Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method.

Chemical Oxygen Demand

Chemical Oxygen Demand Range Low Range: 0 to 150 mg/L
Medium Range: 0 to 1500 mg/L
High Range: 0 to 15000 mg/L
Chemical Oxygen Demand Resolution Low Range and Medium Range: 1 mg/L
High Range: 10 mg/L
Chemical Oxygen Demand Accuracy Low Range: ±5 mg/L ±5% of reading
Medium Range: ±15 mg/L ±4% of reading
High Range: ±150 mg/L ±3% of reading
Chemical Oxygen Demand Method Adaptation of the USEPA 410.4 approved method for the COD determination on surface waters and wastewaters.

Chlorine

Free Chlorine Range 0.00 to 5.00 mg/L
Free Chlorine Resolution 0.01 mg/L
Free Chlorine Accuracy ±0.03 mg/L or ±4 % of reading
Total Chlorine Range 0.00 to 5.00 mg/L
Total Chlorine Resolution 0.01 mg/L
Total Chlorine Accuracy ±0.03 mg/L or ±4 % of reading
Chlorine Method Adaptation of the EPA method 330.5 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 4500-Cl G, DPD method.

Nitrate

Nitrate Range 0.0 to 30.0 mg/L NO3--N
Nitrate Resolution 0.1 mg/L
Nitrate Accuracy ±1.0 mg/L or ±5 % of reading
Nitrate Method Chromotropic acid method.

Nitrogen, Total

Nitrogen, Total Range Low Range: 0.0 to 25.0 mg/L
High Range: 10 to 150 mg/L
Nitrogen, Total Resolution Low Range: 0.1 mg/L
High Range: 1 mg/L
Nitrogen, Total Accuracy Low Range: ±1.0 mg/L or ±5 % of reading
High Range: ±3 mg/L or ±4 % of reading
Nitrogen, Total Method Chromotropic acid method.

Phosphorous, Acid Hydrolyzable

Phosphorous, Acid Hydrolyzable Range 0.00 to 1.60 mg/L
Phosphorous, Acid Hydrolyzable Resolution 0.01 mg/L
Phosphorous, Acid Hydrolyzable Accuracy ±0.05 mg/L or ±5 % of reading
Phosphorous, Acid Hydrolyzable Method Adaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 4500-P E, ascorbic acid method

Phosphorous, Reactive

Phosphorous, Reactive Range Low Range: 0.00 to 1.60 mg/L
High Range: 0.0 to 32.6 mg/L
Phosphorous, Reactive Resolution Low Range: 0.01 mg/L
High Range: 0.1 mg/L
Phosphorous, Reactive Accuracy Low Range: ±0.05 mg/L or ±5 % of reading
High Range: ±0.5 mg/L ±5% of reading
Phosphorous, Reactive Method Low Range: Adaptation of the EPA method 365.2 and Standard Methods, 20th edition, 4500-P E, ascorbic acid method
High Range: Adaptation of the Standard Methods, 20th edition, 4500-P C, vanadomolybdophosphoric acid method.

Phosphorous, Total

Phosphorous, Total Range Low Range: 0.00 to 1.15 mg/L
High Range: 0.0 to 32.6 mg/L
Phosphorous, Total Accuracy Low Range: 0.01 mg/L
High Range: 0.1 mg/L
Phosphorous, Total Resolution Low Range: ±0.05 mg/L ±6% of reading
High Range: ±0.5 mg/L ±5% of reading
Phosphorous, Total Method Low Range: Adaptation of the EPA method 365.2 and Standard Methods, 20th edition, 4500-P E, ascorbic acid method
High Range: Adaptation of the Standard Methods, 20th edition, 4500-P C, vanadomolybdophosphoric acid method

General Meter

Logging Memory 200 samples
Connectivity USB
Display Liquid Crystal Display (LCD)
Power Supply 12 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F); max 90% RH non-condensing
Dimensions 235 x 212 x 143 mm (9.2 x 8.34 x 5.62")
Weight 1.5 Kg (3.3 lb)
Photometer/Colorimeter Light Source tungsten lamps with different narrow band interference filters
Photometer/Colorimeter Light Detector silicon photocell
Ordering Information HI83214 is supplied with sample vials and caps (5), batteries, 12 VDC power supply, and instruction manual.