HI97729-11 is a high quality set of calibration standards that allows users to validate and calibrate the fluoride range found in the compatible HI97759-11 portable photometers that have a 0.00 to 2.00 mg/L (ppm) fluoride range. The HI97729-11 includes one sealed “A” cuvette for zeroing the meter and one “B” cuvette that is used to validate or calibrate. The expected reading from the “B” standard is 1.00 ±0.03 mg/L fluoride @ 25℃. The calibration set is produced in our state-of-the-art manufacturing facility using reagent grade chemicals in a temperature controlled environment.
Supplied with Certificate of Analysis
- Product code
- Lot number
- Expiration date
- Standard value @ 25℃
- Reference meter NIST traceable
Provided storage containers
- Light tight
- Protects from accidental breakage
Validation and calibration of the HI97729 Low Range Fluoride Portable Photometer is quick and easy with the HI97729-11 CAL Check Standards.